REALITi 360 Version Update 11.17.0 is ready for download!

REALITi 360 update 11.17.0

REALITi Version 11.17.0

Latest Improvements

Atlas

We’ve improved CPR functionality and stability on additional monitors:

  • CPR feedback now correctly appears on supported additional monitors when the primary monitor is connected to a CPR device
  • Resolved a crash that occurred when selecting CPR Console as an additional monitor while the primary monitor was connected to a CPR device
  • Fixed crashes during compressions on Schiller Defigard and Physiogard when they were used as additional monitors

We’ve fixed an issue with CPR logging on Atlas Adult Gen 1 and Gen 2 manikins:

  • Recoil depth values in logs now record correctly instead of always showing as “-1”

Corpuls

We’ve fixed issues with prompts and audio in C3:

  • Corrected the DEMAND mode message when ECG electrodes are removed, which now displays “No ECG cable in DEMAND mode” instead of “Pause”
  • Fixed AED text and voice prompts when switching from Monitor mode so they match the pie chart state
  • Restored the “Perform CPR” voice prompt when returning to AED with the pie chart running
  • Adjusted Start Auto Analysis so AED prompts display and sound correctly
  • Resolved an issue where AED voice prompts could still be heard while in Monitor mode

We’ve updated arrow icons and controls for better clarity and consistency in C1:

  • Replaced “Up” and “Down” text buttons with arrow icons on Pacing intensity and frequency screens
  • Improved adjustment behavior for intensity and frequency when using the arrow keys
  • Updated arrow icons in soft key buttons, the Settings menu, and Manual Defib energy selection screens

We’ve improved German translations in Controller:

  • Fixed issues with Advanced ICP Controls, Advanced PAP Controls, and clinical/monitoring terms
  • Added translations for Imported Image and Imported Video labels
  • Corrected casing for the USE OFFLINE button and resolved a translation issue for Urine – Ketones
  • Standardized wording where existing translations were inconsistent

We’ve corrected the default alarm limits for heart rate (HR), oxygen saturation (SpO₂), and pulse rate (PR):

  • C1 and C3 now correctly use their default HR, SpO₂, and PR alarm limits when launched on their own

Facilitator

We’ve updated the default settings for ETCO₂ and respiratory rate in Control and scenarios:

  • ETCO₂ now defaults to 39
  • Respiratory rate (RR) now defaults to 12
  • These values are applied at startup, when connecting a new monitor, and when creating a new scenario
  • The REALITi Welcome Scenario and quick start scenarios have also been updated (this change applies if “copy default scenarios on startup” is selected and the app is restarted; any custom changes to scenarios will be overwritten)

CPR Band

We’ve added logging to improve troubleshooting for the CPR band:

  • Logs are now recorded when using the CPR band, making it easier to identify and resolve issues

Nihon Kohden – EMS1052

We’ve improved translations and language support:

  • Added translations for the 12-Lead ECG button and front panel in Japanese, Spanish, French, German, Portuguese, Italian, Polish, and Romanian
  • Adjusted font size, style, and layout to ensure translated text displays correctly
  • Updated styling for AED and Pacing labels with orange background highlights

We’ve fixed a display issue in Manual mode:

  • The compression graph no longer appears during energy charge when no CPR device is connected

We’ve improved the display of heart rate values in numerics:

  • Fixed an issue where the HR value shrank when highlighted in Large and Normal Numerics
  • Added a restriction so that HR highlighting does not appear when the ECG source is PAD

We’ve made several improvements to the CPR assist feature for clearer feedback during compressions:

  • Bars now show the correct starting compression depth when compressions do not begin from zero
  • Bar visuals updated to improve clarity, including adjusted width/height and the addition of grey bars for better release and compression feedback
  • “CPR Good Compressions” message now displays on a white background
  • Bars turn turquoise when compressions meet the correct depth and rate, while other bars remain red
  • Improved padding for compression values, “CPR-Comp,” and “Tempo /m” labels to ensure consistent spacing

We’ve fixed the ETCO₂ display in Control when connected to EMS1052 to ensure clearer and more consistent feedback:

  • The ETCO₂ eye icon now has only two states: OFF (grey with a line) and ON (blue without a line)
  • Removed the incorrect third “pale” eye state
  • Toggling the ETCO₂ eye now correctly updates both ETCO₂ and RR text visibility together
  • When OFF, the waveform and text are grey and hidden; when ON, they are blue and visible

We’ve updated the UI with new Japanese translations:

  • Corrected mistranslations and missing Japanese labels
  • Fixed line breaks so text displays as intended
  • Adjusted font sizing so Japanese text fits properly without breaking into unnecessary lines
  • Updated age units and other NIBP interval labels
  • Removed extra periods that were incorrectly added to some translations

Lifepak 15

We’ve fixed an issue with cardioversion shocks on the LifePak monitor:

  • Joule settings now remain at the last manually entered value (e.g., 100 J) for subsequent shocks, instead of auto-escalating

Maintenance

We’ve upgraded the Instabug iOS SDK to version 16.0.2 to maintain compatibility with upcoming iOS releases

Zoll X Advanced

We’ve improved patient mode settings:

  • Heart rate and pulse rate alarm parameters now adjust correctly when switching to Pediatric or Neonate modes

Visit the App Store today to update your REALITi 360 system.