REALITi 360 Version 11.23.0: Introducing the Mindray BeneHeart D30

REALITi with Mindray D30

REALITi 360 Version 11.23.0: Introducing the Mindray BeneHeart D30

Now available in the App Store!

Latest Improvements

Alarms

  • Improved alarm behavior on the Generic screen by ensuring alarm limits update correctly when switching patient age profiles, with ETCO₂, respiratory rate, non-invasive blood pressure, and temperature alarms and visual indicators now triggering correctly when values move outside configured limits
  • Corrected Schiller monitor SpO₂ alarm limits so they now follow the selected age profile correctly

Corpuls

  • Improved AED shock recommendation behavior, so shocks are now correctly recommended when ventricular fibrillation (VF) or ventricular tachycardia (VT) rhythms are detected.
  • Improved pacing behavior when ECG waveforms are hidden or disconnected. Pacing now stops appropriately when ECG data is unavailable, and users are prompted to connect therapy electrodes when pacing cannot be displayed.
  • Updated the C3 signal intensity bar to provide a more realistic indication of perfusion level. The bar no longer flashes with every heartbeat and now reflects the selected perfusion setting.

CPR Console

  • Fixed an issue where enabling CPR could display incorrect heart rate values. The configured CPR rate is now shown correctly, the previous heart rate is restored when CPR is disabled, and repeated activation no longer causes heart rate values to increase unexpectedly.
  • Improved readability of CPR timing information by displaying elapsed and hands-off times using a consistent HH:MM:SS format.

Facilitator/Control

  • Improved consistency of blood pressure values, units, alarm limits, and exported recordings across Control, Monitor, Camera, and other connected displays when measurement settings are changed
  • Corrected blood pressure display behavior to ensure systolic and diastolic values remain accurate across connected screens
  • Fixed an issue where non-invasive blood pressure alarm limits were not triggering correctly in some situations

iSimulate Community

  • Improved the reliability of importing, uploading, and downloading Community scenarios
  • Fixed an issue where importing scenarios from REALITi 360 could fail to start
  • Resolved a crash that could occur while downloading Community scenarios, improving overall stability

Lifepak 15

  • Enhanced the AED training workflow with updated voice prompts, on-screen guidance, and improved prompt timing to better match device behavior during training scenarios
  • Added appropriate prompts for non-shockable rhythms, including "Check for pulse" and "If no pulse, start CPR", along with translated text and voice prompts for supported languages
  • Prevented 12-lead ECG acquisition while in AED mode and added a notification explaining when the function is unavailable
  • Improved the 12-lead ECG workflow by retaining patient age and sex between consecutive ECG recordings during the same sessions
  • Updated Manual mode metronome guidance to use spoken ventilation prompts with improved timing for a more realistic training experience
Mindray monitor

Mindray BeneHeart D30

  • Added support for the Mindray BeneHeart D30 monitor

Zoll

  • Corrected CPR depth units and formatting so measurements remain consistent when switching between centimeters and inches
  • Fixed several display issues affecting AED mode, including BVM positioning, CPR display overlap, ventilation information, and Manual Mode exit dialog navigation, resulting in a clearer and more reliable user interface

Misc

  • Improved the confirmation message displayed after submitting a bug report, including clearer guidance and a reminder to connect to the internet so reports can be successfully sent to the support team
CTGI

CTGi Improvements Version 1.0.15-66

We’ve improved the consistency between the fetal heart rate display and the audible heartbeat:

  • Fixed an issue where the audible heartbeat did not match the selected fetal heart rate baseline
  • The displayed baseline and heartbeat sound are now synchronised across supported chart drawing speeds

Visit the App Store today to update your REALITi 360 and CTGi systems