Latest Improvements in REALITi 360 Version 11.15.0

Atlas

  • Fixed manikins on old firmware not getting the correct type
  • Additional checking of manikin types
  • Depth parameter changes with multiple CPR devices attached
  • Fixed timeout
  • Increased timeout to 7 seconds for Atlas Baby, Junior, AdultGen3
  • Fixed pulse level setting not reflecting pulse level
  • CPR artefacts appear on ECG trace when using Atlas
  • SpO2 waveform changes to Poor Perfusion when ventilation or compression is started

Camera

  • Fixed waveforms not running on the camera screen

Corpuls

  • Removed energy escalation protocol settings (C1, C3)
  • Improved C1 invert display
  • Stopwatch responds to scenario pauses and resumes
  • Fixed C1 QRS tone disappearing when in either defibrillator or pacer modes, and C3 QRS tone disappearing when switching from home screen to defib modes
  • Analyze and Charge buttons activate AED and Manual Defibrillator modes
  • Fixed CPR button showing no artefacts on the Pacing rhythm
  • Pacer off prompt appears while pacing is on when changing from Monitor mode to AED or Manual defibrillator
  • Resolved “Connect pacer cable” appearing in C1
  • Removed middle soft key option at the start of AED, including during CPR phase, and second soft key option ‘Manual’ to ‘Energie’ in Manual mode
  • Adjusted the visible soft key options in Manual mode—when charging or fully charged—to display only Energy, Cancel, and Monitor, in both English and German
  • Resolved Metronome setting text in AED and Manual modes
  • Resolved pacing rhythm in Demand mode when only pads are on
  • Extra checks on RA or LL electrodes as part of the pacing rhythm state to ensure it affects pacing rhythm
  • Corrected pacing rhythm not affected by pads when off
  • Improved how the rhythm with pacing reacts when either electrodes or pads are removed and then re-attached
  • Fixed text prompt for Pacing when switching to defib
  • Fixed ‘Switch off Pacer?' transitions when the switch is cancelled while electrodes are off
  • Added translation ‘Schrittmacher ausschalten?’
  • Fixed C1 audio doubling voice prompts
  • Improved C1 Views

Hamilton T1

  • Removed ‘Connect Ventilator’ button
  • Fixed default Delta P value
  • Fixed spontaneous breath marker not being removed at start and end of trace
  • Spont TI is automatically calculated
  • Improved calculations of default TI when changing Adult IBW or Neonatal Weight
  • Improved IBW/weight calculation for the Vt alarm limit
  • Improved the SPONT mode by disabling the option and setting value to 0 in Ventilator settings
  • Fixed Pressure cmH2O alarm limits on high and low for Neonatal and Adult
  • Improved alarm ExpMinVol default limits’ calculation from weight/IBW, Vt and Rate
  • Corrected ExpMinVol low and high limit values for each patient type
  • Fixed ExpMinVol low and high limits where it skips the maximum limit value
  • Fixed fTotal high and low limits, showing Adult limits instead of Neonatal
  • Fixed Vt limit low value skipping value 10 before the Off symbol for Adult
  • Fixed Vt limit high value to show maximum of 3000 for Adult before the Off symbol
  • Fixed Vt limit high value to show maximum of 300 for Neonatal before the Off symbol
  • Fixed Vt limit low value to show maximum of 0.1 for Neonatal before the Off symbol
  • Improved Vt limit low value calculation based on weight or IBW
  • Fixed Alarms limits where the high value can get lower than the low value
  • Improved Rate and TI values
  • Fixed dial subtext
  • Improved TI max dot
  • Added indicator for Nebulizer when it’s active
  • Added message “Nebulizer is active”
  • Created 10-sec timer for the generic message display
  • Added a stop action for the Nebulizer to remove message and turn off light indicator

Facilitator/Control

  • Fixed primary folder appearing empty when it’s not
  • Added multiple lines for scenario names

Generic

  • Fixed Sync not getting disabled after a shock

Nihon Kohden - EMS1052

  • Improved Japanese translations

Philips MRX

  • Display CPR waveform on all leads when CPR is active

Teaching Management System (TMS)

  • Fixed data computation, compression depth guide, and compression target rendering during CPR

Zoll X Advanced

  • Adjusted label sizes so translations fit appropriately

Visit the App Store today to update your REALITi 360 system.